GE
D
IRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
Chapter 3 - System Setup 3-105
NOTE: The IP address must be different from the one used for the Vivid q N system.
8) Type in the Subnet mask 255.255.255.0.
NOTE: The Subnet mask should be the same for both the network computer and the
VividqN.
9) Click OK, close all windows and approve all changes. Return to regular mode.
10) Connect the Vivid q N and network computer LAN connection using a cross over Network cable.
(Normally with RED connector covers on both sides, cable not longer than 100 meters).
The Vivid q N and the network computer are now ready for export, import of files.
NOTE: One can check the connection by pinging to the Vivid q N IP, as follows:
a.) Select <Start>, <Run>
b.) Type CMD, then press <Enter>
c.) Type Ping and IP 10.0.0.1 for Vivid q N then press <Enter>.
If return timing is OK, you have an operating network connection.