GE
D
IRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
3-104 Section 3-10 - Connectivity Setup
a.) Double-click on My Computer icon on desktop.
b.) Double-click on Control Panel icon.
2) Double-click on Network Connections icon.
3) Double-click on Local Area Connection icon.
The Local Area Connection Properties dialog box opens.
4) Select Internet Protocol (TCP/IP).
5) Click the Properties button.
The Internet Protocol (TCP/IP) Properties dialog box opens.
6.) Activate the Use the following IP address check box.
7) Type in the network computer’s IP address 10.0.0.2.
Figure 3-108 Local Area Connection Properties
Figure 3-109 Internet Protocol (TCP/IP) Properties