GE
D
IRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
Chapter 6 - Service Adjustments 6-11
6-1-7-1 Performing the Battery Calibration Procedure
NOTE: It is recommended to commence this procedure towards the end of the working day (when the
battery is fully charged) then allow the procedure to be completed overnight, as outlined in
Figure 6-10.
1) Click on the battery icon at the bottom of the scanning screen
OR
from the message window shown below, click Recalibrate.
The Battery Calibration Wizard (Page 1) opens - Figure 6-11.
2.) Make sure the AC power cable is connected to the Vivid q N, then click Next to commence the first
stage of the process (Full Charge to maximum capacity).
This step will fully recharge the battery and may take up to 4 hours.
Figure 6-11 Battery Calibration Wizard - Page 1