IRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
7-104 Section 7-5 - Monitoring Tests
assistance.
• If errors were identified but the results were very close to the upper or lower limits, perform
calibration, as described in "DC Offset Calibration" on page 7-25, and then repeat the test.
• If errors were identified (results exceeding acceptable upper or lower limits), verify that all
system cables and boards are connected (not misplaced) and functioning properly, as
described in Chapter 5 - Components and Function (Theory).
• If the test fails again, repeat the Fan Speed Test.
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of this board is faulty. Continue trouble-shooting
down to the appropriate board. (If necessary, perform a specific test).
If the problem still persists, contact the GE On-Line Center for assistance.
10.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
11.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.