IRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-105
7-5-1-6 Tx Voltages Test
NOTE: For a description of this test, refer to the information in "Tx Voltages Test" on page 7-14.
1.) In the Diagnostic Test window, trackball to Tx Voltages in the list of Monitoring tests available for
selection from the Diagnostic Test Tree (see Figure 7-72 below) and press Set. The Tx Voltages
Test name is displayed under Execute; the relevant system components to be tested are
highlighted light blue in the Schematic View (AC/DC, Battery, PS), as shown below:
2.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
3.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
4.) Trackball to the Start button and press Set.
The Tx Voltages Test commences. While the test proceeds, the name of the test currently in
progress (Tx Voltages) is displayed in the Execute field and the current Loop # and specific test are
shown below. The messages Init Done and Please Wait are displayed beside the progress bar in
the Status area.
Figure 7-72 Tx Voltages Test