IRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
7-112 Section 7-6 - Performing Back End Diagnostics on the System
7-6-1-2 ECG Test
NOTE: For a description of this test, refer to the information in "ECG Test" on page 7-15.
NOTE: When performing this test, no probe should be connected to the system and nothing should be
touching the probe connectors!
1.) Connect the ECG cable to the input on the ECG port on the right side of the Vivid q N scanner.
2.) Connect the ECG leads to yourself (as the simulated patient).
3.) In the Diagnostic Test window, trackball to ECG Test in the list of BEP tests available for selection
from the Diagnostic Test Tree (see Figure 7-77 below) and press Set. The ECG Test name is
displayed under Execute; the relevant system components to be tested are highlighted light blue in
the Schematic View (BEP, Peripherals, Keyboard), as shown below:
4.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
5.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
6.) Trackball to the Start button and press Set.
The interactive ECG Device Test screen is displayed as shown in Figure 7-78.
Figure 7-77 ECG Test