IRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
7-142 Section 7-7 - Automatic Error Log
1.) Perform step 1 and step 2 on page 7-137.
2.) Click the Extensive Log button (refer to Figure 7-103 on page 7-138).
The Extensive Logging dialog opens - see Figure 7-106, below.
3.) Flag the appropriate checkboxes to activate the categories for which detailed analysis is required
(refer to the recommendations in Table 7-4).
NOTE: The impact of activating a checkbox would be to generate larger, detailed log files.
After usage, make sure to return to the default settings (unflagged) to conserve the size of the
log files. After 7 days, the system will automatically return to the default settings.
4.) When done, click OK.
The Extensive Logging dialog closes.
Figure 7-106 Extensive Logging Dialog