EasyManua.ls Logo

GE Vivid q N - Page 521

GE Vivid q N
698 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE
D
IRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-15
8-2-4-5-5 ATGC Slider Keycap Removal Procedure
NOTE: This section describes the replacement procedure for ATGC Slider keycaps (supplied as part
of the Keyboard Assembly) - Vivid i n systems: Part No. S423357; Vivid q N systems: Part No.
S423358.
1) Carefully place a small flat screw driver between the ATGC slider keycap you wish to remove and
the surface of the keyboard.
2.) Use the screw driver to gently lever off the defective keycap, until it is completely loosened from its
base.
3.) Lift off and remove the defective keycap - see Figure 8-19 below.
8-2-4-5-6 ATGC Slider Keycap Installation Procedure
1) Carefully place a new slider keycap in position on the slider base, taking care to place the plastic
alignment pin in the correct location.
2.) Gently push the slider keycap down until it snaps into position.
8-2-4-5-7 ATGC Rotary Knob Removal Procedure
NOTE: This section describes the replacement procedure for ATGC Rotary knob (supplied as part of
the Keyboard Assembly) - Vivid i n systems: Part No. S423357; Vivid q N systems: Part No.
S423358.
1) Carefully place a small flat screw driver between the ATGC Rotary knob and the surface of the
keyboard - see Figure 8-17 below.
2.) Use the screw driver to gently lever off the defective rotary knob, until it is completely loosened from
its base.
Figure 8-19 ATGC Slider Keycap Replacement

Table of Contents

Related product manuals