EasyManua.ls Logo

GE Vivid q N - Page 525

GE Vivid q N
698 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE
D
IRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-19
8.) Press Config and log on as Adm - see "Log On to the System as ADM" on page 4-10.
9.) Select About, then click on the HwVersion tab.
10.)Verify that the correct ECG/Respiratory module Part # is displayed, as shown in Figure 8-23.
11.)Proceed to perform the following functionality tests:
- "Isolated Patient Lead (Source) Leakage – Lead-to-Ground" on page 10-26
- "Isolated Patient Lead (Source) Leakage – Lead-to-Lead" on page 10-27
- "Isolated Patient Lead (Sink) Leakage - Isolation Test" on page 10-27
- "SafeLock Cart - Grounding Continuity" on page 10-22 - only if the
Vivid q N system is currently mounted on the SafeLock cart
- "SafeLock Cart - Chassis Current Leakage Test" on page 10-24
Figure 8-23 Hardware Version Verification

Table of Contents

Related product manuals