8-24 Section 8-2 - External Component Replacement Procedures
8.) Reconnect the LCD flex cable to the BEP (refer to Figure 8-26 on page 8-22).
9.) Refit the Control Panel and Keyboard, as described in the "Control Panel and Keyboard Installation
Procedure" on page 8-12.
10.)Close and open the lid a few times to ensure proper closure of the LCD Display cover.
11) Reconnect the AC adapter and all cables to the Vivid q N portable ultrasound scanner.
12.)Proceed to perform the following functionality tests:
- "SafeLock Cart - Grounding Continuity" on page 10-22 - only if the Vivid q N system is
currently mounted on the SafeLock cart
- "SafeLock Cart - Chassis Current Leakage Test" on page 10-24