GE
D
IRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
8-30 Section 8-3 - Internal Component Replacement Procedures
8-3-1-4 P&M (Probe and MUX) Board Removal Procedure
1) Remove the bearing handle, as described in the "Bearing Handle Removal Procedure" on page 8-2.
2.) Remove the Control Panel and Keyboard, as described in the "Control Panel and Keyboard
Removal Procedure" on page 8-10.
3.) Loosen and remove the four screws that secure the P&M Board to the RFI Board, as shown in
Figure 8-31.
4.) Loosen the two screws that fasten the P&M Board to the Bottom Assembly, as shown in
Figure 8-32 below.
Figure 8-31 Removing the Screws that Secure the P&M Board to the RFI Board
Figure 8-32 Loosening the Screws from the P&M Board