IRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
8-110 Section 8-5 - Software Loading
14. Type in the option key that you received from the OTR
and click OK.
Note: Do not use option keys from the previous
BT version.
The system will boot-up into Normal scanning mode.
15. Press Config. The Operator Logon screen opens.
16. Type the following to log into the system:
Operator: ADM
Password: ulsadm
Click Logon.
The normal scanning screen is displayed.
17. To verify the software version, press Config and click
About at the bottom of the screen. Select the System
Version tab.
The About screen opens, and from the System Version tab, the installed
software version is displayed on the screen
18. Verify that the software version corresponds with the
System SW version you have installed (check against
the label on the USB Flash Drive) and that it was
correctly installed.