IRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-111
19. Restore Patient Archive and System Configuration
as per user manual.
Note: When performing “Restore”, click Admin at the
bottom of the screen, then click the Restore tab.
Select “System Configuration” and then de-select “GSP
Resources”.
20. Install peripherals.
21. Restore the settings.
Verify proper Connectivity functionality - refer to
"Connectivity Setup" on page 3-85.
22. Proceed to perform the following functionality tests:
"Software Configuration Checks" on page 4-39
"Calibration Tests" on page 7-25
"Full System Test" on page 7-10
Table 8-6 Step-by-Step Vivid q N Software Upgrade/Installation/Reinstallation Procedure
Step Instruction Expected Result
De-select GSP Resources