7-5 FAILSOFT TESTING
7-5.1 System Initiated Failsoft Test
Sequence
Configure the analyzer for trunk mode (paragraph
6-5) and System Initiated Testing (paragraph
6-7).
1. Enter the following parameters:
Sig Type: SYSTEM INIT
ID Disp:
Call Seq: FAILSOFT
2. Enter the
FSTx or channel number
FSRx or channel number (if VHF/UHF)
3. Enter the Port selection and monitor
attenuation. In most cases the port selection
should be RF I/O with 20 dB attenuation.
4. Enter the generator Port selection. In most
cases the port selection should be RF I/O.
5. Press the start test softkey. If the parameters
selected are not valid, the test sequence is
terminated and an error message is
displayed. When the
start test softkey is
pressed, it is replaced with the
stop test
softkey. For a list of error and warning
messages, refer to Appendix E.
The trunking transmits the failsoft word on
the failsoft channel, and nothing on the
control channel. When the radio cannot
detect a control channel, it looks at its
failsoft channel. When the radio decodes the
failsoft word it will unmute and
conventional radio operation is possible.
After the radio has locked on the failsoft
channel, the operator can transmit
modulation from the trunking analyzer to the
radio.
6. Press the stop test softkey to terminate
testing.
132

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the General Dynamics R2670 Series and is the answer not in the manual?

General Dynamics R2670 Series Specifications

General IconGeneral
BrandGeneral Dynamics
ModelR2670 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals