EasyManua.ls Logo

GeoMax Zoom 30 series - Page 180

GeoMax Zoom 30 series
262 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Zoom | 180Calibration
Access 1) Select Tools from the MAIN MENU.
2) Select Calibr. from the TOOLS MENU.
Select:
HA-collimation, or
Vertical Index.
)
The procedures and conditions required to correct line-of-sight and vertical
index errors are the same, therefore the procedure will only be described once.
a) Mechanical vertical axis of the instrument, also called
standing axis
b) Axis perpendicular to the vertical axis. True 90°
c) Vertical angle is reading 90°
d) Vertical index error
)
By determining the vertical index error the elec-
tronic level is adjusted automatically
b
c
d
a
Zoom_012

Table of Contents

Related product manuals