3.2 Working with the Battery
)
Charging / first-time use
The battery must be charged prior to using it for the first time because it
is delivered with an energy content as low as possible.
For new batteries or batteries that have been stored for a long time (>
three months), it is effectual to make only one charge/discharge cycle.
The permissible temperature range for charging is between 0°C to
+40°C/+32°F to +104°F. For optimal charging we recommend charging
the batteries at a low ambient temperature of +10°C to +20°C/+50°F to
+68°F if possible.
It is normal for the battery to become warm during charging. Using the
chargers recommended by GeoMax, it is not possible to charge the battery
if the temperature is too high.
Operation / discharging
The batteries can be operated from -20°C to +50°C/-4°F to +122°F.
Low operating temperatures reduce the capacity that can be drawn; very
high operating temperatures reduce the service life of the battery.
For Li-Ion batteries, a single discharging and charging cycle is sufficient.
We recommend carrying out the process when the battery capacity indi-
cated on the charger or on a GeoMax product deviates significantly form
the actual battery capacity available.