REMOTE CONTROL
181
MEASure2:RESistance?
Returns the 2W resistance measurement on the second display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF),Resolution(<NRf>|
MIN | MAX | DEF)]
Example: MEAS2:RES?
> +1.181372E+6
Returns the 2W measurement.
MEASure2:FRESistance?
Returns the 4W resistance measurement on the second display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF),Resolution(<NRf>|
MIN | MAX | DEF)]
Example: MEAS2:FRES?
> +1.181372E+6
Returns the 4W measurement.
MEASure2:FREQuency?
Returns the frequency measurement on the second display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF),Resolution(<NRf>|
MIN | MAX | DEF)]
Example: MEAS2:FREQ?
> +0.215029E+5
Returns the frequency (21.5 kHz).
MEASure2:PERiod?
Returns the period measurement on the second display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF),Resolution(<NRf>|
MIN | MAX | DEF)]
Example: MEAS2:PER? MAX
Returns the period at the maximum range.
SENSe Commands
[SENSe:]TEMPerature:TCOuple:TYPE
Sets thermocouple type.
Parameter: Type(B | E | J | K | N | R | S | T)
Example: SENS:TEMP:TCO:TYPE J
Sets the thermocouple to type J.
[SENSe:]TEMPerature:TCOuple:TYPE?
Returns the thermocouple type.
Return parameter: B, E, J, K, N, R, S, T