5. Confirm and
go back to the
default display
Press the Enter key to confirm.
Press the Exit key to go back to the
default display.
Set Reference Junction Temperature (T-CUP)
When a thermocouple is connected to the GDM-8261A,
the temperature difference between the thermocouple
lead and the GDM-8261A input terminal should be taken
into account and be cancelled; otherwise an erroneous
temperature might be added.
The terminal temperature is manually defined by the user.
Default value: 23.00
1. Open
reference
junction menu
Press the Shift key, then the TEMP
(Sensor) key. The sensor selection
menu appears on the display.
Press the Left and Right arrow keys
and select T-CUP (thermocouple).
Press Down, Right arrow key and
then Down again. The reference
junction selection menu appears on
the display.