Select Voltage Range
To turn the automatic range selection
On/Off, press the AUTO key.
Press the Up or the Down key to select
the range. The AUTO indicator turns
Off automatically. If the appropriate
range is unknown, select the highest
range.
For more detailed parameters, see the specifications
on page 218.
DC voltages with AC components cannot be accurately
measured if the DC+AC component exceed the ADC
dynamic range for the selected DC range. Any voltage
exceeding the ADC dynamic range will be clipped at the
upper/lower range limit. Under these conditions the range
that is chosen with the Auto range function may be too
small.
A
+2V
0V
-2V
C D
B
E
DC 1V Range
A,B: Input exceeds the ADC
dynamic range.
C,D: The DCV offset causes
the input to exceed the upper
ADC dynamic range.
E: The DCV offset causes the
input to exceed the lower
ADC dynamic range.