BASIC MEASUREMENT
39
Set Continuity Threshold
Background
Continuity threshold defines the maximum resistance
allowed in the DUT when testing the continuity.
Threshold Range
0 ~ 1000Ω, 1Ω resolution, 10Ω default
1. Activate
threshold setting
1. Press the Shift key, the 2nd
key,
the Right key. The measurement
menu appears.
/ EXITSHIFT
2ND
MENU
TRIG
2. Press the Down key, the Right
key, the Enter key. The
continuity threshold setting
appears.
TRIG
AUTO
ENTER
W
2. Edit threshold
1. Move the cursor (the flashing
digit) using the Left/Right key.
HOLD TRIG
2. Change the value using the
Up/Down key.
Range:
1 ~ 1000Ω, 1Ω resolution, default 10Ω
3. Go back to the
default display
Press the Enter key to confirm the
edited threshold. Press the Exit key
to go back to the default display.
AUTO
ENTER
/ EXITSHIFT
www. .com