GDM-8261A User Manual 
114 
1Count# 32 4 5 6 7 8 9 10 11 12
Delay Delay
Timer Timer
Scan
CH
101
CH 
102
CH 
103
CH 
104
CH
101
CH 
102
CH 
103
CH 
104
CH
101
CH 
102
CH 
103
CH 
104
Timer
Delay
Trigger delay time
 
Example: Scan channels 1~4 with a count setting of  12. 
Measures a single channel in the specified 
range (Channel MIN~MAX)at each 
trigger event. Timer settings (page 115) 
apply for each channel. 
1Count#
CH
101
Delay
Timer
2
CH 
102
Delay
Timer
3
Delay
Timer
4
CH 
104
Delay
Timer
5
CH 
101
Delay
Timer
6
CH 
102
Step
CH 
103
12
Timer
Timer
Delay
 
Example: Step through channel 1~4 with a count of 12.   
Selects just one channel and continuously 
measures it. 
Ensure the scanner has been installed before trying to 
configure the scanner (page 104). 
1.  Press the Shift key, the 2nd key 
(MENU), the Left key. The Scan 
menu appears.   
2.  Press the Down key. The Simple 
Scan menu appears. 
www. .com