FAQ
I pressed the EXIT key but cannot get out of Scanner mode.
The GDM-8261A performance does not match the specifications.
I pressed the EXIT key but cannot get out of Scanner mode.
Press the Exit key, followed by the ACV (Scan) or DCV
(Step) key.
The GDM-8261A performance does not match the specifications.
Make sure the device is powered On for at least 1 hour.
This is necessary to stabilize the unit to match the
specifications.
How can I achieve the fastest measurement speed
To achieve the fastest measurement speed, the DMM
must use the internal trigger and must also be used in
remote control mode using the USB interface. The
following settings should also be set remotely:
1. The measurement mode and/or range settings
applicable for the measurement. For example:
DCI:
CONF:CURR:DC 1 (see page 176)
SENS:CURR:DC:NPLC 0.025 (see page 194)
DCV:
CONF:VOLT:DC 1 (see page 173)
SENS:VOLT:DC:NPLC 0.025 (see page 194)
2W:
CONF:RES 1000 (see page 176)
SENS:RES:NPLC 0.025 (see page 194)
4W:
CONF:FRES 1000 (see page 176)
SENS:FRES:NPLC 0.025 (see page 195)