REMOTE CONTROL
183
[SENSe:]TEMPerature:FRTD:TYPE
Sets the 4W RTD sensor type.
Parameter: Type(PT100 | D100 | F100 | PT385 | PT3916 | USER)
Example: SENS:TEMP:FRTD:TYPE PT100
Sets the 4W RTD sensor to PT100
[SENSe:]TEMPerature:FRTD:TYPE?
Returns the 4W RTD sensor type.
Return parameter: PT100, D100, F100, PT385, PT3916, USER
[SENSe:]TEMPerature:FRTD:ALPHa
Sets the 4W RTD Alpha coefficient.
Parameter: <NRf> (0~10)
Example: SENS:TEMP:FRTD:ALPH 0.00385
[SENSe:]TEMPerature:FRTD:ALPHa?
Returns the 4W RTD Alpha coefficient.
[SENSe:]TEMPerature:FRTD:BETA
Sets the 4W RTD BETA coefficient.
Parameter: <NRf> (0~10)
Example: SENS:TEMP:FRTD:BETA 0.00495
[SENSe:]TEMPerature:FRTD:BETA?
Returns the 4W RTD BETA coefficient.
[SENSe:]TEMPerature:FRTD:DELTa
Sets the 4W RTD DELTa coefficient.
Parameter: <NRf> (0~10)
Example: SENS:TEMP:FRTD:DELT 0.0000568
[SENSe:]TEMPerature:FRTD:DELTa?
Returns the 4W RTD DELTa coefficient.
[SENSe:]DETector:RATE
Sets the detection rate (sample rate)
Parameter: RATE(S | M | F)
Example: SENS:DET:RATE S
Sets the rate to slow (S).
[SENSe:]DETector:RATE?
Returns the sample rate.
Return parameter: SLOW, MID, FAST