[ 1 ] Specifications are for 1-hour warm-up at 6 1⁄2 digits, Slow ACfilter,
sinewave input.
[ 2 ] Relative to calibration standards.
[ 3 ] 20% overrange on all ranges, except 750 Vac, 10A range.
[ 4 ] Specifications are for sinewave input >5% of range. For inputs from
1% to 5% of range and <50 kHz, add 0.1% of range additional error. For 50
kHz to 100 kHz, add 0.13% of range.
[ 5 ] 750 Vac range limited to 100 kHz
[ 6 ] Typically 30% of reading error at 1 MHz.
[ 7 ] For frequencies below 100 Hz, slow AC filter specified for sinewave
input only.
[ 8 ] Additional settling delay required when input dc level varies.
[ 9 ] 0˚C~18˚C, 28˚C~55˚C