ADVANCED MEASUREMENT
67
5. Result
High
If the 2nd display shows
High, the result is above
the High limit.
Digital I/O: FAIL Out (Pin 6) and
HIGH Limit FAIL Out (Pin 7) are
activated.
Low
If the 2nd display shows
Low, the result is below
the Low limit.
Digital I/O: FAIL Out (Pin 6) and LOW
Limit FAIL Out (Pin 8) are activated.
Pass
If the 2nd display shows
Pass, the result is staying
between the High and the
Low limit.
Digital I/O: PASS Out (Pin 5) is
activated.
Digital I/O
The Compare measurement result
comes out from the rear panel
Digital I/O terminal. For the
terminal details, see page 124.
Deactivate
Compare
measurement
To cancel the Compare
measurement, press the Shift key
followed by the Hold (Comp) key, or
simply activate another
measurement.
/ EXITSHIFT
HOLD
COMP