CONFigure:CURRent:AC
Sets measurement to AC Current on the first display and
specifies the range.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: CONF:CURR:AC 10e-2
Sets the measurement mode to ACI with a 100mA range.
CONFigure:CURRent:DCAC
Sets measurement to DC+AC Current on the first display and
specifies the range.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: CONF:CURR:DCAC 10e-2
Sets the measurement mode to DC+AC Current with a 100mA
range.
CONFigure:RESistance
Sets measurement to 2W Resistance on the first display and
specifies range.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: CONF:RES 10e3
Sets the range to 10kΩ.
CONFigure:FRESistance
Sets measurement to 4W Resistance on the first display and
specifies range.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: CONF:FRES 10e3
Sets the range to 10kΩ.