CONFigure2:RESistance
Sets measurement to 2W Resistance on the second display and
specifies the range.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: CONF2:RES 10e3
Sets the range to 10kΩ on the second display.
CONFigure2:FRESistance
Sets measurement to 4W Resistance on the second display and
specifies the range.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: CONF2:FRES 10e3
Sets the range to 10kΩ on the second display.
CONFigure2:FREQuency
Sets measurement to Frequency on the second display and
specifies the range.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: CONF2:FREQ MAX
Sets the frequency measurement range to max on the second
display.
CONFigure2:PERiod
Sets measurement to Period on the second display and specifies
the range.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: CONF2:PER
Sets the DMM to period measurement using the previous range
on the second display.
CONFigure2:OFF
Turns the second display function off.
Parameter: None.