SYSTem:RWLock
Enables remote control and disables local control (front panel
control). Once this command has been issued, pressing the 2ND
or local buttons will not return the user to local control. The only
way to return to local mode is to issue the SYSTem:LOCal
command.
IEEE 488.2 Common Commands
*CLS
Clears the Event Status register (Output Queue, Operation Event
Status, Questionable Event Status, Standard Event Status)
*ESE?
Returns the ESER (Event Status Enable Register) contents.
Example: *ESE?
>130
Returns 130. ESER=10000010
*ESE
Sets the ESER contents.
Parameter: <NR1> (0~255)
Example: *ESE 65
Sets the ESER to 01000001
*ESR?
Returns SESR (Standard Event Status Register) contents.
Example: *ESR?
>198
Returns 198. SESR=11000110