OPERATION
45
Diode Test
The diode test checks the forward bias characteristics of a diode by
running a constant forward bias current of approximately 1mA
through the DUT.
Set to Diode
Measurement
1. Press the
/
key once to activate diode
measurement.
Note: pressing the
/
key twice will activate
the continuity measurement instead.
2. The mode will switch to Diode mode
immediately, as shown below.
Display
Diode
state
Diode function
indicator
CVCCSWEEPFILTRCLSTOCOMP2ND
SHIFTERRLIMITAUTOMATH4W2WDCAC
mdB
%
AVmm
HzWkM
WSFC
RMTMAXMINRELHOLDSMFAUTO4W2WDC
Connection
Connect the test lead between the VΩ
terminal and COM terminal; Anode-V,
Cathode-COM. The display updates the
reading.
1000V
750V
500Vpk
INPUT
FUSE
1.25A/1000V
V
W
SENSE
W
4W
COM
200Vpk
H I
LO
10A 1A
CAT II 600V