APPENDIX
141
Replacing the AC Source Fuse
Fuse Ratings
Type
Rating
Size
0.125AT
100VAC, 120VAC
5mm X 20mm
0.063AT
220VAC, 240VAC
5mm X 20mm
Note
Only replace the fuse with a fuse of the correct type
and rating.
Steps
1. Turn the DMM off and take out the power cord.
2. Remove the fuse socket using a flathead
screwdriver.
3. Remove the fuse in the holder and replace with
the correct type and rating.
4. Ensure the correct line voltage is lined up with
the arrow on the fuse holder. Insert the fuse
socket.