COMMAND OVERVIEW
119
MEASure2:FRESistance?
Returns the 4W resistance measurement on the second display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: MEAS2:FRES?
> +1.00023E+03
Returns the 4W measurement.
MEASure2:FREQuency?
Returns the frequency measurement on the second display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: MEAS2:FREQ?
> +1.01122E+03
Returns the frequency (1.01122kHz).
MEASure2:PERiod?
Returns the period measurement on the second display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: MEAS2:PER? MAX
Returns the period at the maximum range.
SENSe Commands
[SENSe:]TEMPerature:TCOuple:TYPE
Sets thermocouple type.
Parameter: Type(J | K | T)
Example: SENS:TEMP:TCO:TYPE J
Sets the thermocouple to type J.
[SENSe:]TEMPerature:TCOuple:TYPE?
Returns the thermocouple type.
Return parameter: J, K, T