GDM-8351 User Manual
30
Basic Measurement Overview
Refresh Rate
Background
The refresh rate defines how frequently the
DMM captures and updates measurement data.
A faster refresh rate yields a lower accuracy. A
slower refresh rate yields a higher accuracy.
Consider these tradeoffs when selecting the
refresh rate.
For further details, please see the specifications.
Refresh rate
(Reading/S)
Function
S
M
F
Continuity/Diode
10
40
320
DCV/DCI
10
40
320
ACV/ACI
10
40
320
Frequency/Period
1
9.8
83
Temperature
10
40
320
Resistance
10
40
320
Capacitance
2
2
2
Steps
1. Press the left or right arrow keys to change the
refresh rate.
2. The refresh rate will be
shown at the top of the
display.
F M S
Note
The refresh rate cannot be set for capacitance
measurement.