OPERATION
33
Select the Voltage Range
The voltage range can be set automatically or manually.
To turn the automatic range selection On/Off,
press the AUTO key.
Press the Up or the Down key to select the
range. The AUTO indicator turns Off
automatically. If the appropriate range is
unknown, select the highest range.
Selectable Voltage
Ranges
For further details, please see the specifications on
page 145.
DC voltages with AC components cannot be
accurately measured if the DC+AC component
exceeds the dynamic range for the selected DC
range. Any voltage exceeding the dynamic range
will be clipped at the upper/lower range limit.
Under these conditions the range that is chosen
with the Auto range function may be too small.