MSO-2000E & 2000EA User Manual
150
Timeout
Triggers when the signal stays high, low or either
for a designated amount of time. The trigger level
determines when a signal is high or low.
A
B
C
A
Trigger level threshold
B
Timer
C
Triggering point
Bus
Triggers on SPI, UART, I2C, Parallel, CAN or LIN
bus events.
Logic
Triggers on specified logic levels or for specified
clock edge. Logic trigger is only available for
Digital channels.
Trigger: types and sources
Sources versus
types
Trigger
types
Trigger sources
Analog
Digital
CH1 ~ CH4
EXT*
AC Line
D0 ~ D15
Edge
Delay
Pulse Width
Video
Pulse &
Runt
Rise & Fall
(Slope)
Timeout
Bus
**
**
Logic