MSO-2000E & 2000EA User Manual
184
11. Press Trigger When from the side
menu to choose the triggering
condition for the data.
When
=, , <, >, ≤, ≥
12. The oscilloscope will now trigger when the
specified bus data matches the Trigger When
conditions.
Parallel Bus Trigger
The parallel bus trigger conditions can be set at any time after the
bus setting has been set to parallel. The parallel bus can be set up to
trigger on a specified data pattern.
Panel Operation
1. Set the Bus to parallel in the bus
menu.
Page 135
2. Press the Trigger Menu key.
Menu
3. Press Type from the bottom menu.
4. Select Others Bus from
the side menu. The Bus
indicator appears at the
bottom of the display.
From left: Bus trigger, Data source
5. Press Data from the bottom menu.