ADVANCED CONFIGURATION
185
6. Press Data from the side menu to
edit the triggering data.
To edit the data, use the Variable
knob to highlight a binary or hex
digit and press Select. Use the
Variable knob to choose a value for
the digit and press Select to
confirm.
VARIABLE
Binary
0,1,X (don’t care)
Hex
0~F, X (don’t care)
7. The oscilloscope will now trigger when the
specified data appears on the bus.
Using the Logic Trigger
Background
The digital channels can be set up to trigger on
specified logic levels and for a specified clock edge.
For example the digital channels can be set to trigger
on the rising edge of a clock signal when bit 1 of a
digital channel is high and all other channels are
ignored.
Panel Operation
1. Press the Trigger Menu key.
Menu
2. Press Type from the bottom menu.
3. Select Others Logic from
the side menu. The Logic
indicator appears at the
bottom of the display.

Table of Contents