WGM - 28 November 2007 - Edition 01
PAT700 Total Organic Carbon Analyzer – Bottle Mode Page 99
Anatel Operator Manual
Run now icon immediately draws a water sample from the online source into the
excursion bottle and opens the bottle filling dialog box. See Figure 6-27.
Done icon returns you to the home screen. When an excursion condition occurs,
the analyzer performs an excursion sample capture. Following the excursion, the
analyzer always returns to the home screen.
Cancel icon returns you to the unload/replace bottles dialog box. All changes made
in any of the dialog boxes are lost.
Figure 6-27 Bottle filling dialog box
4) When an excursion occurs, an excursion sample captured dialog box opens, and a
validation occurs. See Figure 6-28.
Following the excursion and validation, a validation results dialog box opens that
contains the blank and 500 ppb sucrose data. The next arrow on the validation
results dialog box takes you to the validation summary dialog box.
The analyzer writes the date, time, TOC value, analyzer serial number, status,
oxidation curve filename, conductivity, and temperature information to the RFID tag.
Figure 6-28 Excursion sample captured dialog box