131 UnableToStart4to20mAOutputs Analyzer cannot start 4-20 mA output test.
132 UnableToStartPumpTest Analyzer cannot start pump test.
133 UnableToStopPumpTest Analyzer cannot stop pump test.
134 UnableToCalibratePump Analyzer cannot calibrate pump.
135 UnableToExportDataLog Analyzer cannot export data log.
136 UnableToExportAuditTrail Analyzer cannot export audit trail.
137 UnableToAddUser Analyzer cannot add a user.
138 UnableToEditUser Analyzer cannot edit a user.
139 UnableToDeleteUser Analyzer cannot edit a user.
141 UnableToAcknowledgeAlarms Analyzer cannot acknowledge alarm.
142 UnableToDeleteAlarms Analyzer cannot delete alarms.
143 UnableToStartSelfClean Analyzer cannot start self clean.
144 UnableToStopSelfClean Analyzer cannot stop self clean.
145 UnableToStartOnlineManualSample Analyzer cannot start online manual sample.
146 UnableToStopOnlineManualSample Analyzer cannot stop online manual sample.
147 UnableToRestoreFactoryDefaults Analyzer cannot restore factory defaults.
148 UnableToStartVoltageTest Analyzer cannot start voltage test.
149 UnableToStartPlumbingTest Analyzer cannot start plumbing test.
150 UnableToPrintDataLog Analyzer cannot print data log.
151 LampExtinct • Analyzer detects that current lamp has failed.
• Replace UV lamp.
153 Bottle1MarkedAsUsed • Bottle installed in location 1 is marked as used.
• Replace bottle with a valid standard.
154 Bottle2MarkedAsUsed • Bottle installed in location 2 is marked as used.
• Replace bottle with a valid standard.
155 Bottle3MarkedAsUsed • Bottle installed in location 3 is marked as used.
• Replace bottle with a valid standard.
156 Bottle4MarkedAsUsed • Bottle installed in location 4 is marked as used.
• Replace bottle with a valid standard.
157 ExcursionBottle1NotEmpty • Excursion bottle 1 is not empty.
• Remove excursion bottle 1 and replace with empty bottle.
158 ExcursionBottle2NotEmpty • Excursion bottle 2 is not empty.
• Remove excursion bottle 2 and replace with empty bottle.
Table 7-3 Warning alarm messages