Mechanical installation
26 A0608-14.0 HBM: public T10F
WARNING
The screw heads (Z), see Fig. 6.4, must not touch the
adapter flange.
With alternating load: Use a screw locking device
(e.g. LOCTITE no. 242) to glue the screws into the
counter thread to exclude prestressing loss due to screw
slackening.
3. Before the final tightening of the screws, rotate the
torque flange on the centering device until all screw
heads are positioned approximately centrally in the
through‐holes of the connection element. The screw
heads must not touch the walls of the through‐holes in
the adapter flange!
4. Fasten all screws with the specified tightening torque
(see Fig. 6.4).
5. For further mounting of the shaft train, there are eight
tapped holes on the adapter flange. Also use screws
of property class 10.9 (or 12.9) and fasten with the
torque specified in Fig. 6.4.
Important
With alternating loads, use a screw locking device to
cement the connecting screws into place! Guard against
contamination from varnish fragments.
The maximum screw‐in depth as per Fig. 6.4 must be
complied with! Otherwise, significant measurement errors
may result from torque shunts or the transducer may be
damaged.