H E Y E R P a s i t h e c
72 / 106 HEYER Pasithec, Operator’s manual Rev. 0.2 Draft - 12/09
Step 2
Turn the knob to select “Calibrate” submenu.
Press the knob and the cursor appears on the O
2
Sensor Calibration option.
Step 3
Press the knob and a new window appears right side.
Perform the operation in accordance with the prompt on the screen.
Disassemble O
2
sensor from the Y-piece and place it in the air for more than 3 minutes.