5.1 Triggering Measurement Start and Stop
104
Using Logic Triggering
Logic triggering is available when Logic is selected for pulse input
channels (p. 61).
Triggering is controlled by the signal state and combination of
logic input signal channels.
Select a trigger pattern (1, 0, or X) and AND/OR combining logic
so that triggering occurs when the combined criteria are satisfied.
Make trigger settings on the [CH] screen.
See:"Key Setting Procedure" (p .101)
T T
High
Low
T T T
High
Low
1 Select the trigger signal combining logic.
(Select [OR] or [AND] as the start/stop trigger type.)
2 Select the triggering state (pattern) for each
pulse channel.
(This setting applies to the pulse input channels se-
lected for trigge
r logic.)
Setting options: (
: default setting)
OR
Triggering occurs when the state of any one
of the trigger signals matches the set pattern.
AND Triggering occurs only when the states of all
of the trigger signals match the set pattern.
Setting options: ( : default setting)
X
Signal is ignored.
0 Triggering can occur when the signal
switches from High to Low.
1 Triggering can occur when the signal
switches from Low to High.