3.6 Scaling Settings (as needed)
69
3
3
Chapter 3 Settings
Use the scaling function to convert input volt-
age to the physical units of the measurement
p
arameter for display, such to convert voltage
input for display as electrical current.
Converted values can be disp
layed in fixed or
floating-point notation.
Make these settings on the [CH] screen.
See:"Key Setting Procedure" (p .65)
3.6 Scaling Settings (as needed)
Normal Display
(Off)
Scaling Enabled
A
mV
The Setting screen cannot be dis-
played while measuring.
Scaling Voltage, Temperature, Humidity, Resistance, and Revolution
Measurements
1 Select the desired numerical notation.
2 Select the scaling method.
:
3 Enter the [Unit] name on the comment entry
screen. (Up to 7 characters)
See: About Comments (p. 71)
4 Ratio: Enter the conversion ratio and offset.
2-pt: Enter the values at two points before and
af
ter conversion.
Move the cursor to the numerical value and press EN-
TER to
display the numerical setting dialog.
See: "Numerical Value Entry" (p .67)
Setting options:(
: default setting)
Dec Displays converted values using fixed-point
notation.
Exp Displays converted values using floating-
point notation.
Off
Scaling is disabled.
Setting options:( : default setting)
Ratio
Convert by specifying the ratio of physical
units per volt of input signal (conversion ra-
tio), offset, and the name of the units.
2-pt Convert by specifying the input signal volt-
age at two points, the physical unit values at
those
points, and the name of the units.
12 3
Conversion Ratio Setting
2-Point Setting
2-Point Setting
With [Cond] set to [Exp]
With [Cond] set to [Dec]
4
Conversion Ratio Setting
A value measured on CH1 of UNIT1 (Analog channel CH1-1) can be set as the scaling
value (Inter-Channel Compensation function) (p. 79).