3.3 Input Channel Settings
60
Make these channel settings for resistance measurement.
Make these settings on the [CH] screen.
See: "Key Setting Procedure" (p .53)
Resistance Measurement Settings
1 Select the input unit (UNIT1 to 4) and channel
(1-1 to 4-15), and check the box to enable the
channel.
2 Select [Resist].
3 Select the appropriate measurement ampli-
tude range.
Setting options:( : default setting)
10, 20, 100, 200
1
23
: On
: Off
For inductive loads such as coil windings, the slow response time may pre-
vent correct resistance measurement. In this case, select a recording time
longer
than 5 s.
When a measured value far exceeds the range limit (range label)
"OVER" or "-OVER" is displayed in place of numerical/cursor values, and the following values
are applied to calculations and data storage.
Input type Input range +OVER value -OVER value
Resistance
10 16.3835 -16.384
20 32.767 -32.768
100 163.835 -163.84
200 327.67 -327.68