3.5 Waveform Display Settings (as needed)
68
Displayed waveforms can be expanded or compressed along
the horizontal axis. Shorter setting values correspond to greater
magnification. Make these settings on the [Setting] screen.
See:"Time Setting by Key Operations" (p .50)
Settings can be changed during and after measurement on the
[Wave+Value] display of the Waveform/Numerical screen(p. 89).
The [Time/DIV] range depends on the [Interval] setting.
Horizontal axis display options are affected when the recording interval is changed after measuring, be-
cause the options for expanding (or compressing) the displa
yed axis depend on the recording interval used
when measuring. When using the Logger Utility program with a LAN connection, recording intervals from 100
ms to 2 s are not available, regardless of auto save settings (5 s is automatically selected).
Setting the Display Time Base (horizontal axis magnification)
1 div
Select the time per division.
Setting options:( : default setting)
100ms
, 200ms, 500ms, 1s, 2s, 5s, 10s, 20s, 30s, 1min,
2min, 5min, 10min, 20min, 30min, 1h, 2h, 5h, 10h, 12h,
1day [/DIV]
: Selectable, ×: Not selectable, : Not selectable while measuring
Horizon-
tal axis
Recording interval
[ms] [s] [min] [h]
10 20 50 100 200 500 1 2 5 10 20 30 1 2 5 10 20 30 1
100ms
× × × × × × × × × × × × × × × ×
200ms
× × × × × × × × × × × × × × ×
500ms
× × × × × × × × × × × × × × ×
1s
× × × × × × × × × × × × ×
2s
× × × × × × × × × × × ×
5s
× × × × × × × × × × × ×
10s
× × × × × × × × × ×
20s
× × × × × × × × ×
30s
× × × × × × × × ×
1min
× × × × × × ×
2min
× × × × × ×
5min
× × × × × ×
10min
× × × ×
20min
× × ×
30min
× × ×
1h
×
2h
5h
10h
12h
1day
When auto save is enabled for waveforms
100, 200 and 500 ms settings are automatically changed to 1 s when measurement starts, and
the shorter values cannot be selected while measuring.