7.2 Numerical Value Calculation Expressions
148
7.2 Numerical Value Calculation Expressions
Average
Obtains the average value of waveform data.
AVE: A
verage value
n: D
ata count
di: Data on channel number i
Peak Value
(P-P)
Obtains the value of the difference (peak-to-peak
value) between maximum and minimum values of
waveform data.
Maximum
Value
Obtains the maximum value of waveform
data.
Minimum
Value
Obtains the minimum value of waveform
data.
Time to max-
imum value
Acquires the time (s) elapsed from the start of re-
cording to maximum value. When there are two or
more maxima, this value is the first to occur.
Time to mini-
mum value
Acquires the time (s) elapsed from the start of re-
cording to minimum value. When there are two or
more minima, this value is the first to occur.
Interval Cal-
culation
(Auto Calc
only, P.144 )
When [S
plit Save] is enabled ([On] or [Ref
Time]), calculation results are saved at the speci-
fied interval.
When [O
n]:
Set the [S
plit Length]. Calculation results for the
specified time span are automatically saved.
When [R
ef Time]:
Set the [R
ef Time] and [Split Length]. Calcula-
tion results starting from the
reference time are au-
tomatically saved for the specified time span.
AVE
1
n
-- -
di
i1
=
n
=
Maximum
value
Minimum
value
Peak value
Maximum
value
Minimum
value
Maximum
value
Time to maximum value
Start of Measurement
Time to minimum value
Minimum
value
Start of Measurement
Split Time
Start of
Measurement
End of
Measurement
When
[On]
Reference
Time
Start of
Measurement
End of
Measurement
When
[Ref
Time]
Split
Length

Table of Contents