2.4 Making Connections
43
3
2
Chapter 2 Measurement Preparations
< Connecting to the external control terminals >
Required items: phillips screwdriver (for M3 screws), measurement leads
External Control (using TRIG OUT and EXT TRIG)
1 Lift the external control termi-
nal block cover.
2 Loosen the terminal block
screws, insert the bare ends of
the leads, and tighten the
screws.
3 Connect the trigger signal
input and output.
4 Replace the cover on the exter-
nal control terminal block.
Connect the '+' wire to the
TRIG.OUT (or EXT.TRIG) terminal,
and the '-' wire to a GND terminal.
Connect the TRIG.OUT terminal a
trigger input on another device, and
connect the EXT.TRIG (input) termi-
nal to the trigger output another
device (as needed).
Five GND terminals are provided.
Connections can be made to any
GND terminal.
GND
2
3