5.4 Periodic (Timer) Measurements
115
3
5
Chapter 5 Specifying Criteria for Measurements
Make these settings to record at a specific time. Recording can be set to
repeat at specific intervals between the set
start and stop times.
Before setting, confirm that the HiLogge
r clock is set to the correct
time. If not, reset the clock on the [S
ystem] screen (p. 158).
5.4 Periodic (Timer) Measurements
Select the [Setting]
Screen.
Move to a setting item.
Apply
Open the setting options
for the item to be set.
Select from the listed op-
tions
1
2
Change 1's
digit
Change 10's
digit
Key Setting Procedure
1 Set the [Timer] to [On]. (default setting: Off)
2 Enable (set to [On]) the desired timer setting items.
3 Set the year, month, day, hour, minute and second for the start or stop time.
To enter the current time as the start or stop time, select the clock mark ( ) and press
ENTER.
4 (When [Interval] is set to [On])
Set the amount of time (days, hours, mi
nutes and seconds) between
recording repetitions.
5 Press START.
[Wait Timer]
will be displayed at the bottom of the waveform screen.
Setting options:
Start Set the recording start time.
Stop Set the recording stop time.
Inter-
val
Set the recording repetition interval.
This setting is enabled on when [Rep
eat] is set to [On] (p. 52).
1
2 3
4
Setting Procedure