6.2 Saving Data
127
3
Chapter 6 Saving & Loading Data
6
During real-time saving, removable storage can be replaced without interrupting mea-
surement. This procedure describes replacing a USB flash drive.
Replacing Removable Storage During Real-Time Saving
Move the cursor to [EJECT].
Execute
Insert a formatted removable
storage.
Measurement data acquired while the de-
vice was removed is automatically written.
See: Formatting: (p. 47)
Insertion: (p. 45)
Remove the USB flash drive (p. 46).
Right bottom of the screen
Move the cursor to
[Unmount USB memory].
Execute
Move the cursor to [Yes].
Execute
Execute
A "Replaceable" message appears.
Bottom of the screen
When both a CF card and USB flash drive are in-
stalled and either is removed, the save destina-
tion automatically switches to the remaining
media.
• If a storage device is not replaced within two minutes after selecting "Yes" in
the "Do you want to remove the USB memory/CF card?" confirmation dialog,
data may overflow internal memory and be lost.
• If a storage device is replaced during real-time saving, data is saved to a
new file.
• If measurement ends while the removable storage is ejected, even if a
removal storage is then inserted, the remaining data is lost. In that case,
when connected to a LAN, data can be read from the HiLogger via the HTTP
function. See "10.4 Remote Measurement with the HTTP Server Function"
(p. 188).