3.3 Input Channel Settings
62
Configure these settings for each pulse channel receiving input from a pulse output
device such as a watt-hour or flow meter. Make these settings on the [CH] screen.
See: "Key Setting Procedure" (p .53)
Integration (Count) Measurement Settings
1 Select the input unit (PLS&ALM) and channel
(P1 to P8), and check the box to enable the
channel.
2 Select [Count].
3 Select the integration mode (Count Mo).
4 Select the basis of measurement (Slope).
5 Select the High/Low reference (threshold) val-
ues.
6 Select a filter setting.
Setting options:( : default setting)
Add
Counts the number of pulses acquired from
the start of measurement.
Inst Counts the number of pulses acquired dur-
ing each recording interval. The pulse count
i
s reset at the beginning of each recording
interval.
Setting options:( : default setting)
Counts the number of L - H transitions.
Counts the number of H - L transitions.
Setting options:( : default setting)
1V
Judge 1.0 V or higher as HIGH, and 0 to
0.5 V as LOW.
4V Judge 4.0 V or higher as HIGH, and 0 to
1.5 V as LOW.
Setting options:( : default setting)
Off
Filter is disabled.
On Filter is enabled. Prevents incorrect counts
due to relay contact chatter (noise).
Totalized
[Add]
Instantaneous
[Inst]
0 V
L
H
1
: On
: Off
2
3
4
5
6
• Use the scaling function to display integrated pulse counts as the cor-
responding physical units (such as Wh or VA) of the measurement
p
arameter (p. 69).
• The counter saturates at 1,073,741,823 counts. If you expect to
exceed this count, we recommend measuring with the Integration
Mode (Count Mo) set to Instantaneous (Inst), and summing later with
a spreadsheet program.