5.5 Measurement with Trigger and Timer Functions
118
About Dead time
After a recording length has elapsed, some “dead time” is needed for internal pro-
cessing before recording can resume. No measurement occurs during the dead
time. So to avoid information loss, split data into files at a specified interval by
enabling continuous recording ([Repeat]), and selecting [Split Save] for auto sav-
ing.
Off
Start/
St
op
Start/
Stop
A start trigger oc-
curs after the timer
start
time, and a
stop trigger occurs
before the timer
stop time
On Stop Off
Cont.
Tr
ig-
ger
Timer
Repeat: Off Repeat: On
Press START Timer start/stop time Trigger criteria match
T
Press STOP
End of
Measurement
Start of
Measurement
Measuring
Recording time
T T
End of
Measurement
Start of
Measurement
Measuring
Recording
time
Dead time
Recording
time
T T T T
End of
Measurement
Start of
Measurement
Measuring
T
End of
Measurement
Start of
Measurement
Measuring
Dead time
T T
Dead time
Measuring